The pulse duration is controlled by Dazzler and measured by Wizzler.
Dazzler - is a turn-key ultrafast pulse shaping system, performing simultaneous and independent spectral phase and amplitude programming of ultrafast laser pulses.
Dazzler systems inserted in CPA systems can pre-compensate gain narrowing and to compensate high order phase distortions, thus enabling the generation of shorter and cleaner ultrafast pulses.
Wizzler - Single shot, high contrast pulse measurement.
Spectral detection band edges: 560 - 1040 nm.
Pulse duration range: 20 - 100 fs (FWHM values for FTL Gaussian pulses).
Temporal measurement window: ± 400 fs.
Temporal measurement dynamic: > 40 dB.
Required pulse energy: 2 - 20 μJ.
The combination of the Dazzler pulse shaping capabilities with the Wizzler high dynamic range measurement provides the most powerful compression optimization tool.
×